How to go MAD at CHESS

                                            home
 
   Getting Beam Time       chess : cuinfo : beam status : ithaca : weather  

 

 
About MacCHESS
MacCHESS People
Publications
Acknowledging MacCHESS
 
Beam lines
Beam Status
A1 station
F1 station
F2 station
F3 station
 
Data collection
BioSAXS
Microcrystallography
High Pressure Cooling
MacCHESS automounter
Crystallography software
Data backup
How to
 
Technical information (primarily for staff)
Troubleshooting guide
Crystal centering problems
F1 Automounter
Improvements in adxv
Saving Images and Video
LED lighting
Small computers
 
 
Contents
Is MAD Phasing Feasible for this System?
Before Going to CHESS
Energy Scans
Energy Selection
Planning Data Collection
Geometry
CCD
Timing
Monitoring Data Collection
Summary
References
How to start the experiment on F2
 
  So, you are a macromolecular crystallographer and have crystals of an interesting protein. You know you can get a set of native diffraction data, but phasing is a problem. You can't do molecular replacement because this is a brand new protein, and you haven't been able to find any usable isomorphous heavy atom derivatives. However, you can make a non-isomorphous derivative, or you know you can get cloned protein incorporating selenomethionine - looks like a case for MAD (multiple wavelength anomalous diffraction).
 
First question: is MAD phasing feasible for this system?
  In order for MAD phasing to be successful, the signal from the anomalous scatterers must be sufficiently large relative to the total scattering from anomalous plus non-anomalous scatterers. The anomalous signal from each heavy atom depends on what that atom is, and you must also consider the available x-ray flux at the wavelength of the appropriate absorption edge. A database of absorption edges is available, courtesy of Ethan Merritt.
 
  At the CHESS F2 station, with the standard Si (111) monochromator crystals, the range of possible energies is 7.9 - 14 keV (1.57 - 0.89 A). This allows use of the K edges of elements 28 - 36 (including Cu, Se, Br, etc.) or the L-III edges of elements 67 - 84 (including Hg, Pt, Au, some lanthanides, etc.). The total signal depends on the number of (well-ordered) anomalous scatterers as well as on their nature. As an example, consider a 100 residue protein containing 2 Se atoms. The total number of
electrons is approximately 5200. Of this, the two selenium atoms contribute 68. The single-wavelength anomalous (Bijvoet) signal depends on the imaginary component of the scattering factor, f'', while the multiple-wavelength (dispersive) signal depends on the difference in scattering factors at the two wavelengths, f'(1)-f'(2). These factors vary with the environment of the Se atoms, but typical values are at most 3.7 and 6.6 electrons, respectively [1], assuming suitable selection of wavelengths. The resulting ratio of anomalous signal to total structure factor may then be calculated according to the method of Hendrickson [2], and is at most 4.0% for the Bijvoet and 3.5% for the dispersive case.
 
  Table 1 gives the maximum number of residues per heavy atom for a reasonable signal (i.e. 3% Bijvoet differences and 2% dispersive differences) from some representative anomalous scatterers, using f' and f'' values determined in protein crystals. Your crystals may have a larger f'', and you may be able to get useful information from a smaller signal, if you are very careful in collecting data, but
this table is a good place to start in deciding whether a MAD experiment is likely to succeed.
 
  Table 1  Maximum Amino Acid Residues per Heavy Atom for Good MAD Signal
Heavy atom Absorption edge Maximum f" Maximum delta (f') Max. no residues
per heavy atom
Cu 1.381 ? (K) 4.2 5.0 90
Se 0.980 ? (K) 3.7 6.6 90
Br 0.920 ? (K) 3.4 5.5 75
Hg 1.009 ? (L-III) 10.2 13.3 640
Ho 1.536 ? (L-III) 25.7 19.9 1430
         

 
The last column of this table was calculated using an inversion of the equation given by Hendrickson in Reference 2 for calculating anomalous signal ratios. The f" and delta (f') values are taken from References 3 (Cu), 1 (Se), 2 (Br), 4 (Hg), and 5 (Ho).
 
  Because the MAD signal is so small, it is especially important to minimize errors in intensity measurements. This means: use your best crystals, freeze them, and be careful in selecting data collection parameters. It is worth spending considerable time optimizing crystallization and freezing conditions (before coming to CHESS, of course). During data collection, select each crystal carefully and optimize exposure time, oscillation range, etc. for it - see information on "Efficient collection of oscillation data" for hints.
 
  Return to Contents
 
Before going to CHESS
  Having decided to attempt a MAD experiment at CHESS, follow the usual procedures to request beam time. As with any experiment involving heavy atoms, you must consider the total amount and form of heavy atom compounds to be brought to CHESS. If heavy atoms are only present in prederivatized crystals, the experiment is non-hazardous - just check the appropriate place on the proposal form. If you are bringing unfrozen crystals in a soak solution containing heavy atoms, you will need to include a "Heavy Atom Compounds Declaration" describing the type and amount of heavy atom compounds that you will be bringing. The small amount of heavy atoms present in the crystals themselves plus a few milliliters of millimolar soak solutions poses no hazard and no administrative complications beyond supplying the Declaration. If you expect to bring concentrated stock solutions of toxic compounds (not recommended but occasionally necessary), you will also need to complete a "Hazardous Materials Declaration". See the heavy atom guidelines for more detail.
 
  You will need a sample containing your heavy atom for taking reference energy scans (see below). CHESS has reference samples for a number of commonly used heavy atoms on hand; check with the staff to see if they have (or can make) one for your experiment. If necessary, make up a standard and bring it along. A simple and satisfactory method is to sandwich a pinch of powdered material (e.g. mercuric acetate for a mercury sample) between two pieces of x-ray transparent sticky tape
(Kapton tape is good), and mount the sandwich in a slide mount. The area covered by the powder should be at least a centimeter or so square, so that it is easy to be sure the x-ray beam is passing through it.
 
  If you are bringing a reference sample containing toxic material, such as a mercury compound, be sure to include it on the Heavy Atom Compounds and Hazardous Materials Declarations. If you think you might need to make up another sample while at CHESS and are bringing a bottle of material to do so, include that also. When you are at CHESS, handle any reference samples you bring with care - label them, try not to spill them, know how to clean them up safely if you do spill them, and don't bring any more than necessary in the first place.
  Return to Contents
 
Energy scans
  Energy scans are essential to select the energies (wavelengths) at which to collect data and to check for drifting of the monochromator during the experiment. These are taken by scanning the monochromator through the appropriate energy range and monitoring the x-ray fluorescence from the sample. CHESS staff will set the station monochromator at the absorption edge of the element you are using; scans may then be easily performed in the vicinity of this energy. Reference samples
containing high concentrations of commonly used elements are provided (see above for how to make your own if you are using something unusual). The reference sample is mounted towards the front of the hutch; the "refscan" command moves it into the beam, scans the energy over a range of about 40 eV (specified using the "madsetup" command), lists and plots the resulting fluorescence spectrum, moves the sample out of the beam, and resets the energy to its initial value. Reference
scans are run periodically during the experiment to check, and if necessary recalibrate, the monochromator. In most cases, you will also want to take an energy scan on your crystal; this can be a little more tricky.

 
  Ideally, you take an energy scan on the crystal you will be using for data collection. However, if this crystal is too small to give a usable fluorescence signal, another (bigger) crystal, or multiple crystals, may be used. In order to get the best signal-to-noise ratio, the fluorescence detector should be placed at 90 degrees to the beam and in the plane of the ring, i.e. pointing along the spindle axis. As normally configured, the cold stream nozzle comes in at an angle and does not interfere with the fluorescence detector, so you may do a crystal scan on a frozen crystal. If you prefer to have the cold stream coming in along the spindle axis, you will have to temporarily remove the cooling nozzle and use unfrozen crystal(s) for the crystal scan. Unfrozen crystals are perfectly satisfactory for this purpose, as the environments of the anomalously scattering atoms change very little on freezing.

Here is a good place to use those junky crystals that are not suitable for diffraction - pack a bunch of them into a capillary and do a crystal scan with them. If possible, tilt the capillary relative to the spindle axis, so that the fluorescent x-rays don't have to go through too much glass. Once you have a sample in the beam, the "xtal_scan" command runs the scan and outputs the results. Panels a - c of the figure below show scans across the Se edge. In panel a is a reference scan taken on a foil. The scan in panel b was taken using a medium-sized crystal with the detector poorly positioned; the panel c scan used a large crystal sample with the detector correctly placed. Panel d of the figure shows a reference scan for Hg. It is typical of this element that the L-III edge is rather broad, with no "white line".
   
  Experimenters have also found empirically that there is very little variation in the edge from one sample to another, so this is a case where you may get away without a crystal scan at all, if you are sure there is mercury in the crystal and are pressed for time.
 
   
Return to Contents
 
Energy selection
  Select your desired wavelengths from the crystal scan. For a 4-energy MAD experiment, use:
(1) a point slightly below the edge,
(2) the inflection point of the edge
(3) the top of the "white line" peak,
(4) a point about 200 eV above the edge

For a typical 3-energy experiment, omit energy (1). For some elements, such as Hg, it may be better to omit energy (3) instead. In case of severe time limitations, useful data may be obtained using just two energies, e.g. (2) and (3) or (1) and (4).
 

  An alternative to a full MAD experiment is the single wavelength anomalous diffraction (SAD) technique; this method uses only the Bijvoet differences collected at a single wavelength. For SAD, the preferred energy is (3), at which f" is largest. Look at the f' and f'' curves for your element (see Ref. 4 for how to calculate these curves from an x-ray absorption spectrum) to decide what selection of energies will
give you the best signal in case you can't collect all the data you would like to. It is generally preferable to have complete data at fewer wavelengths rather than partial data at more wavelengths.
   
  Note that a reasonable anomalous signal can be obtained over a range of energies above the absorption edge, and this range can be quite large for some elements. Hence SAD data may be collected at A1 for Se-containing samples (station tuned 40 eV above the Se edge) and at F1 for Br-containing samples (station tuned 40 eV above the Br edge). At F2, a SAD experiment may be possible even if the station energy cannot be tuned as low as the desired absorption edge. Consult CHESS staff about this possibility if you are working with an element such as I or Xe.
 
  Return to Contents
 
Planning data collection proper
  Geometry Once you have selected the wavelengths for data collection, get a good crystal mounted and determine the appropriate oscillation range and exposure time. Typical values are 1 degree and 1 minute, but these depend heavily on your particular crystal. Decide whether you will be using "mirror, "inverse beam", or no special geometry. In "mirror" geometry, anomalous pairs appear on each image, and in the "inverse beam" case anomalous mates appear on pairs of images
collected with spindle angles differing by 180 degrees. Mirror geometry requires the crystal to be oriented with a mirror plane perpendicular to the spindle axis. At CHESS, the only means for getting this orientation is the use of the arcs on the goniometer head, so that it may not be possible to get a mirror where you would
like it. There can also be a problem in scaling images from a precisely oriented crystal together, particularly in the case of a monoclinic crystal of low mosaicity. Nonetheless, mirror geometry is clearly preferable in terms of reducing the number of exposures and insuring good scaling between anomalous pairs, and should generally be used when possible, especially when another data set is available for scaling purposes. For inverse beam collection, software settings allow collection of wedges of data separated by 180 degrees. It is also possible to collect data in a single sweep, from a randomly oriented crystal, as for a monochromatic experiment; this can be perfectly satisfactory for a robust crystal of relatively high symmetry.

 
  CCD detector MAD users at CHESS usually use a CCD detector. Considerations in your planning due to the detector include:
    You will need to take a background image every time you change the exposure time. This doesn't take long, but is important.
    The readout time, including overhead, is about 3 seconds for the ADSC Q-210 detector (at F2 and A1) and about 1 second for the ADSC Q-270 detector (at F1); you must include this time when calculating how many exposures you can get per hour.
    The detector area is limited. To get high resolution data, it may be necessary to offset the detector, so that the direct beam position is no longer in the center of the detector. For mirror geometry, a vertical offset, i.e. perpendicular to the spindle, will still allow recording of anomalous pairs on each image; a horizontal offset will eliminate some pairs. In the inverse beam case, corresponding reflections appear in opposite quadrants on the paired images, and any offset will eliminate some of them.
 
  Timing Data collection proper consists of a series of energy changes, oscillation exposures, and crystal rotations. A sequence for a 3-wavelength experiment using inverse beam geometry, for example, might be:
    go to wavelength 1, take 5 degrees of data, rotate crystal -5 degrees,
    go to wavelength 2, take 5 degrees of data, rotate crystal -5 degrees,
    go to wavelength 3, take 5 degrees of data, rotate crystal (180 - 5 = 175) degrees,
    go to wavelength 1, take 5 degrees of data, rotate crystal -5 degrees,
    go to wavelength 2, take 5 degrees of data, rotate crystal -5 degrees,
    go to wavelength 3, take 5 degrees of data, rotate crystal -180 degrees.
 
    Repeat until data set is complete, crystal dies, or fill ends.
 
    The major contributor to data collection time is the actual exposure time (at least at F2), which can vary from a few seconds to a few minutes, depending on the crystal quality. Use a series of snapshots to determine the optimal exposure time. Knowing the detector readout time, the time to change wavelengths (about 3 seconds), and
the time to rotate the spindle 180 degrees (about 2 seconds), it is then straightforward to calculate how much data you can collect per hour. Adding in the time to refill CESR (about 10 minutes every 2 - 4 hours), will you have time to get your data? If not, consider reducing the number of wavelengths or the resolution, or plan to come back another time.

 
    Return to Contents
 
Monitoring data collection
  You need to monitor the data collection - sorry, it's not safe to set up an 8-hour data collection and take the day off. It is particularly important to keep an eye on the images being produced to be sure that the crystal is still diffracting. It is also highly desirable to pause data collection whenever the ring is being refilled. The data collection software will halt collection if collimator counts fall too low, but even if this does not happen the beam motion that occurs during filling makes images collected at that time of dubious quality. If collection is halted automatically, you will need to restart it once beam returns. Reference scans should be taken at least once per fill, to check for any monochromator drifting. Once some data have been collected, start processing them to be sure that the quality is sufficient to be worth continuing with the current crystal. Processing as you go will also let you know when enough data have been collected, and may even let you solve a structure at the beamline.
 
  Don't forget to keep up with data backup, so as not to be left with several hours worth of file transfer to do at the end of your run.
 
  Return to Contents
 
Summary of what you need to go MAD successfully
  A suitable molecular system with enough expected signal for phasing.
  Good freezable crystals, and (preferably) a suitable sample for a crystal scan.
  Beam time at CHESS.
  Well-thought out strategy for data collection, with contingency plans.
  Careful experimental technique, with due consideration for safety.
  Good bookkeeping skills.
  A sense of humor to get you through the inevitable equipment failures.
 
References
  1 Yang, W., Hendrickson, W. A., Crouch, R. J., and Satow, Y. "Structure of RNase H Phased at 2A Resolution by MAD Analysis of the Selenomethionyl Protein", Science, 249, 1398-1405 (1990)
 
  2 Hendrickson, W. A. "Determination of Macromolecular Structures from Anomalous Diffraction of Synchrotron Radiation", Science 254, 51-58 (1991)
 
  3 Walter, R. L., Ealick, S. E., Friedman, A. M., Blake, R.C., II, Proctor, P., and Shoham, M. "Multiple Wavelength Anomalous Diffraction (MAD) Crystal Structure of Rusticyanin: a Highly Oxidizing Cupredoxin with Extreme Acid Stability", J. Mol. Biol. 263, 730-51 (1996).
 
  4 Tesmer, J. J. G., Stemmler, T. L., Penner-Hahn, J. E., Davisson, V. J., and Smith, J. L. "Preliminary X-ray Analysis of Escherichia coli GMP Synthetase: Determination of Anomalous Scattering Factors for a Cysteinyl Mercury Derivative", Proteins: Structure, Function, Genetics 18, 394-403 (1994)
 
  5 Weis, W. I., Kahn, R., Drickamer, K., and Hendrickson, W. A. "Structure of the Calcium-Dependent Lectin Domain from a Rat Mannose-Binding Protein Determined by MAD Phasing", Science 254, 1608-15 (1991).
 
    Return to Contents

 

Marian Szebenyi

 

 

 

  Last updated March, 2012
   
CHESS NIH NIGMS